Honeywell Sensing and Control     61
SC 2000, SC2001, SC3004
 Rev. G,  008-0608-00
KNOWN POINT x/y Menu Items
This enters the engineering units for the known-load calibration 
points. These points must match the actual loads that you will 
apply to the instrument during calibration. The number of calibra-
tions points depends on the CALIBRATION TYPE. 
When using the 2-Point Known Load Calibration type, the follow-
ing menu items are available:
• “KNOWN POINT 1/2”: point 1 of 2, usually 0% of the transduc-
er’s full-scale capacity.
• “KNOWN POINT 2/2”: point 2 of 2, usually 100% of the trans-
ducer’s full-scale capacity.
When using the 3-Point Known Load Calibration type, the follow-
ing menu items are available:
• “KNOWN POINT 1/3”: point 1 of 3, usually 0% of the transduc-
er’s full-scale capacity.
• “KNOWN POINT 2/3”: point 2 of 3, usually 50% of the transduc-
er’s full-scale capacity.
• “KNOWN POINT 3/3”: point 3 of 3, usually 100% of the trans-
ducer’s full-scale capacity.
When using the 5-Point Known Load Calibration type, the follow-
ing menu items are available:
• “KNOWN POINT 1/5”: point 1 of 5, usually 0% of the transduc-
er’s full-scale capacity.
• “KNOWN POINT 2/5”: point 2 of 5, usually 25% of the transduc-
er’s full-scale capacity.
• “KNOWN POINT 3/5”: point 3 of 5, usually 50% of the transduc-
er’s full-scale capacity.
• “KNOWN POINT 4/5”: point 4 of 5, usually 75% of the transduc-
er’s full-scale capacity.
• “KNOWN POINT 5/5”: point 5 of 5, usually 100% of the trans-
ducer’s full-scale capacity.
NOTICE
To insure both correct operation of the transducer and ap-
plication of the load, the instrument expects the voltage or 
current applied at each known-load point to be increasing. 
For example, the load applied at Known-Load Point 2/2 must 
cause the transducer to produce a more positive voltage or 
current than at Known-Load Point 1/2.
These menu items are only available with the Known-Load Cali-
bration types.
12.6.6 CALIBRATE Menu Item
This menu item performs a calibration according to what was en-
tered in the CALIBRATION TYPE and CALIBRATION DATA menu items.
NOTICE
Before performing a calibration,
• thetransducermustbeconnectedtotheinstrument,
• thechannel’sExcitationandSignalJumpersmustbeset,
• theCALIBRATION TYPE must be selected (see “CALIBRATION 
TYPE Menu Item” on page 59), and 
• theCALIBRATION DATA must be entered (see “CALIBRATION 
DATA Sub-Menu” on page 60).
NOTICE
For maximum accuracy, allow at least twenty minutes of 
warm-up with the excitation voltage applied to the transducer 
before calibration.
If the CALIBRATION TYPE is Shunt Calibration...
• ThedisplaywillreadDOING SHUNT CAL, and prompt you to 
APPLY 00000 UNIT (where 00000 and UNIT are previously 
entered as ZERO SCALE VALUE and DISPLAY. UNITS. When you 
have applied this load to the transducer, press [ENTER].  
• ThedisplaywillthenreadWORKING and calibration will take 
place. Upon completion, the display will indicate DONE and 
the instrument will return to the RUN mode.
If the CALIBRATION TYPE is 2-Point Known Load Calibration...
• ThedisplaywillreadDOING 2POINT CAL, and prompt you to 
APPLY 00000 UNIT (where 00000 and UNIT are previously 
entered as KNOWN POINT 1/2 and DISPLAY. UNITS). When you 
have applied this load to the transducer, press [ENTER]. 
• ThedisplaywillreadWORKING, then APPLY 01000 UNIT (where 
01000 and UNIT are previously entered as KNOWN POINT 2/2 
and DISPLAY. UNITS). When you have applied this load to the 
transducer, press [ENTER].
• Uponcompletion,thedisplaywillindicateDONE and the 
instrument will return to the RUN mode.