Honeywell Sensing and Control     5
SC 2000, SC2001, SC3004
 Rev. G,  008-0608-00
2.4 Use the SETUP Menus to Enter 
Transducer Information
You can skip this step if:
• YouareusingatransducerequippedwithSignature Calibra-
tion, or
• Thetransducerandinstrumentwerepurchasedtogetherand
set up by Honeywell.
Otherwise, you must enter information about your transducer into 
the SETUP menu of the channel to which it is connected. See the 
appropriate chapter of this manual for that card type. 
2.5 Calibrate the Transducers to Their 
Channels
You can skip this step if:
• YouareusingatransducerequippedwithSignature Calibra-
tion, or
• Thetransducerandinstrumentwerepurchasedtogetherand
set up by Honeywell.
Otherwise, you must choose a calibration method (e.g. Shunt 
Calibration, mV/V Calibration, or Known Load Calibration) ap-
propriate for your application and use the SETUP menu for that 
channel to calibrate the channel to the transducer. See the appro-
priate chapter of this manual for that card type. 
2.6 The SC Series Instrument is Ready 
for Use
See “RUN Mode” on page 6 for information on how to operate the 
instrument while it is in the RUN mode.
Apply some test stimulus on the transducer to observe changes 
in the display.