Revision 12 ST 700 SmartLine Pressure Transmitters User’s Manual Page 161
Table 34 ST 700 Basic Transmitter: Standard Display Menu
English Language
Menu
Submenu/Selection/
Value Entry
Description
Action
Pressure Units
[ 1UNITS]
atm
bar
ftH2O68F
gf/cm2
inH2O39F
inH2O60F
inH2O68F
inHg0C
kgf/cm2
kPa
mbar
mmH2O4C
mmH2O68F
mmHg0C
MPa
Pa
psi
Torr
mH2O4C
cmH2O4C
Choose appropriate
engineering units from list for
calibration and writing
LRV/URV.
This selection determines the
units of the values shown on
the following menu items:
Enter LRV
Enter URV
Set LRV
Set URV
Zero Correct
LRV Correct
URV Correct
Cal Low Point
Cal High Point
Cal Low Correct
Cal High Correct
Barometric Correct
For calibration, this parameter
allows the user to match the
value displayed on the menus
to the units supported by the
user’s calibration equipment.
Press to
enter menu
selection.
Menu to
select from
list
to enter
Enter LRV
[ 2ENTLRV]
Enter URV
[ 3ENTURV]
###. ##
###. ##
The limits are:
2X the Lower Range Limit
(LRL) of the Meter body and
2X the Upper Range Limit
(URL) of the Meter body.
The LRV/URV value will be
available in Pressure Units
(Simple Display pressure unit)
Press to
enter menu
selection.
Menu to
select
number.
to enter
and shift to
the next digit
to the right.
SET Zero
[4SETZRO]
Zero Trim
perform an input calibration
correction by applying process
input at zero.
Press to
enter menu
selection.
Press to
initiate
action.