Probe attenuation (or gain) exceeds calibration limits
The automatic probe cal cannot generate the cal factors to compensate for
the gain or attenuation of the probe. The attenuation factor limits are
80 dB
to
120 dB..
Setup defaulted
This normally occurs when you ask for setup to be defaulted. At power on,
the setups are defaulted if a memory error was detected or had occurred
prior to power down. The setup is also defaulted when new system firmware
is installed or when the memory is declassified in the utility menu.
System firmware error occurred
This is a defect diagnostic that should be reported to Hewlett-Packard
service. Normally, the address where the error occurred is also displayed.
Record the address to help in servicing the instrument. The instrument is
still operable.
System load failed: System firmware is unchanged
This can be caused by either a corrupt disk file or a hardware failure. Try
another system disk file to narrow the problem to the file or the hardware. If
it is a hardware failure, call Hewlett-Packard for service.
Messages
10-6