Recall
Recall sets up the front panel by recalling a front-panel setup from a selected
setup memory. The message "Setup recalled" is displayed at the top-left
corner of the screen indicating the setup was recalled.
If a setup is recalled and you changed plug-ins, the oscilloscope matches the
recalled channel settings as close as possible to capabilities of the new
plug-in.
Default Setup
The default setup key returns the oscilloscope to its default settings. The
default settings place the oscilloscope in a known operating condition. This
known operating condition is used as a starting point in the service
procedures. You may find it helpful to use this known operating condition as
a starting point when setting up the oscilloscope to view signals.
Table 14-1
Default setups
Global
Run/Stop Run
Menu Time base
Time base
Scale
20
µ
s/div
Windowing Disabled
Position 0 s Window scale
10
µ
s
Reference Center Window position 0 s
Trigger
Mode Edge Glitch polarity Negative
Level 0 V Glitch width 20 ns
Sweep Auto Pattern When entered
Hysteresis Normal Pattern definition All don’t care
Holdoff 140 ns Pattern GT 20 ns
Edge source First channel Pattern LT 50 ns
Glitch source First channel Pattern range 20 ns to 50 ns
Slope Positive Delay time 30 ns
Setup Menu
Recall
14–4