General Characteristics
These general characteristics apply to the HP 54700-Series mainframes.
Environmental
Conditions
The instruments meet Hewlett-Packard’s environmental specifications
(section 750) for class B-1 products with exceptions as described for
temperature and condensation. Contact your local HP field engineer for
complete details.
Temperature
Operating
10
°
C to +40
°
C (50
°
F to +104
°
F)
Non-operating
40
°
C to +70
°
C (
40
°
F to +158
°
F)
Humidity
Operating
up to 95% relative humidity (non-condensing) at +40
°
C
(+104
°
F)
Non-operating
up to 90% relative humidity at +65
°
C (+149
°
F)
Altitude
Operating
up to 4,600 meters (15,000 ft)
Non-operating
up to 15,300 meters (50,000 ft).
Vibration
Operating
Random vibration 5-500 Hz, 10 minutes per axis, 0.3 g
rms
Non-operating
Random vibration 5-500 Hz, 10 minute per axis, 2.41
g
rms
; Resonant search, 5 to 500 Hz swept sine, 1 Octave/minute sweep
rate, 0.75g, 5 minute resonant dwell at 4 resonances per axis.