Display Update
Maximum Display Update Rate:
550 Kpixels/s
HP-IB Transfer
Maximum HP-IB Transfer Rate:
500 Kbytes/s
Throughput
This throughput data was taken in the real-time sampling mode (250 MSa/s)
with 512-point records onscreen, no measurements (waveforms/s only), no
interpolation, fast draw mode, infinite persistence, markers off, math off, and
one channel acquisition. Vp-p does not require threshold detection, but
period does.
Throughput Waveforms/second Measurements/second
Measurement Vp-p Period
Front-Panel Capture
and Transfer Rate
>170 >44 >39
HP-IB Capture and
Transfer Rate
>50 >33 >31
Front Panel
Calibrator
ac Output
A 2 kHz or 500 kHz square wave with approximately 1.6 ns
transitions can be used for probe compensation and system calibration.
The waveform levels are 0 V and 1 V when terminated into 50
Ω
.
Specifications and Characteristics
Characteristics
16–8