HP 54710A - Page 306

HP 54710A
412 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Frame
The information within the box titled "Frame" is the configuration of the
mainframe.
Model Number
Indicates the model number of the mainframe.
Serial Number
Indicates the serial number of the mainframe.
Calibration Memory
Indicates the position of the Frame cal switch on
the rear panel. This switch is for the memory in the mainframe that
contains the calibration factors from the last mainframe calibration. This
switch is normally set to the protected position; before calibrating the
mainframe, set the switch to the unprotected position. Make sure you
that you return it to the protected position after completing the
calibration.
Power On Test
Indicates if the power-on self-tests passed or failed.
These power-on tests verify that the six boards are in the mainframe
(five boards in the HP 54710 mainframe). A "Failed" message indicates
that at least one board is loose or defective. The power-up routine is
kept simple to enable the oscilloscope to power up quickly. You can
execute a more complete set of self-tests from the self-test menu.
Firmware Revision
Indicates the revision number of the system
firmware that is installed in the flash ROM memory.
Utility Menu
System Configuration
19–5