EasyManua.ls Logo

HP 54710A - Page 355

HP 54710A
412 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Store screen
Store screen determines what the oscilloscope does to the data on the screen
on a failure. The screen data is a pixel dump of the entire screen area. The
choices are off, pixel memory, printer, disk. The screen image allows you to
see what the display looked like at the time of the failure , but you cannot
rescale it later like you can with waveform memories or functions. If the help
menu is displayed on the screen at the time of the failure, then the help menu
is the screen image that is stored.
Off
Off means that the an image copy of the entire screen area is not
stored, printed, or retained.
Pixel memory
Pixel memory stores an image copy of the entire screen
area to the pixel memory; the same pixel memory that is accessed by the
Waveform menu. Additional failures are added to the composite data
stored in the pixel memory. The only way to clear the pixel memory is
from the Waveform menu.
The value of the pixel memory selection is that you can save the waveform
area from multiple failures to the pixel memory and have a composite view of
the failures.
See also
Chapter 20, "Waveform menu," for additional information on the pixel
memory.
Limit Test Menu
Fail Action
22–12

Table of Contents

Related product manuals