EasyManuals Logo

HP 8712ET User Manual

HP 8712ET
507 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #277 background imageLoading...
Page #277 background image
ES User’s Guide 6-31
Calibrating for Increased Measurement Accuracy
Calibration Kits
How to download a calibration kit ASCII file. After creating a
calibration kit ASCII file, you can download the program by following
these steps:
NOTE Calibration kit definitions must be in DOS format. LIF format is not
supported for cal kit definitions.
1. Insert the disk with the cal standard definitions into the analyzer's
built-in disk drive.
2. Press and note the first word of the second line on
the screen. If it is not "INT:\" (for internal disk), press
.
3. Highlight the file (CALKIT, in the case of the Example Programs
Disk).
4. Press and wait for the prompt:
Loaded cal kit information for 4 calibration standards
Recall of cal kit from CALKIT complete
5. Press , and use the
front panel knob or the keys to highlight the desired
port in the display’s table. Next, select the softkey that corresponds to
the User Cal Kit where the ASCII program stored the calibration kit
(default is User Cal Kit A).
NOTE Cal kit coefficients are displayed in the cal kit block of the operating
parameters screen; press
; press a total of four times.
Step 3: Verify
Performance
Once a measurement calibration has been generated with a user-defined
calibration kit, its performance should be checked before making device
measurements. To check the accuracy that can be obtained with the
user-defined calibration kit, a device with a well-defined frequency
response (preferably unlike any of the standards used) should be
measured. The verification device must not be one of the calibration
standards: measurement of one of these standards is merely a measure
of repeatability.
To achieve more complete verification of a particular measurement
calibration, accurately known verification standards with a diverse
magnitude and phase response should be used. NIST-traceable
standards (such as HP verification standards) are recommended to
achieve verifiable measurement accuracy.
SAVE RECALL
Select Disk
Internal 3.5” Disk
Prior Menu
Recall State
CAL
More Cal
Cal Kit
More Cal Kits
SYSTEM OPTIONS
Operating Parameters
Next Screen

Table of Contents

Other manuals for HP 8712ET

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 8712ET and is the answer not in the manual?

HP 8712ET Specifications

General IconGeneral
BrandHP
Model8712ET
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals