9-6 ES User’s Guide
Specifications
Table 9-1 System Performance, Corrected, 50Ω, 2-Port Calibration (continued)
a. These specifications apply for measurements made using “fine” bandwidth, no averaging, and at an environmental tem-
perature of 25°±5° C, with less than 1° C deviation from the calibration temperature.
b. For transmission measurements, the effect of crosstalk is disregarded and S
12
= S
21
for S
21
<1.0, S
12
=
1
/ S
21
for S
21
> 1.0.
HP 8712ES/8714ES (50 Ω)
HP 85038A (7-16, 50 Ω) Cal Kit, User 2-Port Calibration
Description
Specification
a
(in dB)
300 kHz
to 1.3 GHz
1.3 GHz
to 3 GHz
Directivity 40 40
Source Match 37 37
Load Match 40 40
Reflection Tracking
±0.1 ±0.09
Transmission Tracking
±0.054 ±0.063
Transmission Uncertainty (Specification)
a
,
b
Reflection Uncertainty (Specification)
a