ES User’s Guide 11-3
Factory Preset State and Memory Allocation
Factory Preset and Peripheral States
SOURCE Key Settings
1. HP 8712ES
2. HP 8714ES
3. Analyzers with Option 100 only
4. Preset power level is user-defined by using the
key. The factory default is 0 dBm.
Start frequency 0.3 MHz
Stop frequency
1
1300 MHz
Stop frequency
2
3000 MHz
Low pass/Band pass
3
Low pass
Band pass max span
1,3
1299.700 MHz
Band pass max span
2,3
2999.700 MHz
Frequency resolution kHz
Sweep type Frequency
Alternate sweep Off
Sweep time Auto (fastest possible)
Power level preset power level
4
RF power On
Power sweep start power 0.0 dBm
Power sweep stop power 1.0 dBm
FREQ
SWEEP
POWER