8 - 46
PN 074-519-P1C
3000 Micro GC Operating Manual
17 Set gas delivery pressure and check for leaks. See Table 8-5.
8.5.3.2 Before Turning Off the Carrier Gas
The internal carrier gas cylinder carries sufficient gas for 35 to 40 hours, under
typical operating conditions. Leave the carrier gas ON between analyses to protect
the column(s) and detector(s).
If it is necessary to turn off the carrier gas between analyses, refer to section 8.1.2
on page 8-3 for instructions on shutting down 3000 Micro GC.
CAUTION
Abruptly stopping the carrier gas flow to a hot column
and filament may result in damage to Portable
3000 Micro GC.
Table 8-5 Carrier gas delivery pressure
Carrier Gas Required Delivery Pressure
helium* 552 +/- 14 kPa (80 +/- 2 PSI)
argon 552 +/- 14 kPa (80 +/- 2 PSI)
nitrogen 552 +/- 14 kPa (80 +/- 2 PSI)
* Required for checkout