3.5 Sample Conditioners
CAUTION
3000 Micro GC is a gas-only analyzer!
Extreme caution must be taken to ensure that liquid and
solid particles are not introduced into 3000 Micro GC.
A 10 micron sample inlet filter (PN G2801-60900) is provided in the 3000 Micro GC
ship kit and must be connected to instrument sample inlet at installation. In
addition, a variety of sample conditioners are available to ensure sample entering
the instrument is in gas phase and that the sample gas pressure is below 25 PSI.
Available sample conditioners:
Genie Filter Assembly
Pressure Reducer
Pressure Reducer and Genie Filter Assembly
Heated Regulator
Heated Vaporizer for Liquefied Petroleum Gas (LPG)
For more detailed description of each sample conditioner, see section 5.1, Sample
Conditioners, on page 5-1.