7 - 49
PN 074-519-P1C
3000 Micro GC Operating Manual
7.14 EZ IQ Operation to Activate Remote Start/Cancel Function
To use the remote connector function in EZ IQ:
1 Open EZ IQ.
2 Click Method >> Instrument Setup to display the Instrument Setup window.
3 On the Trigger tab, select Type: External. See Figure 7-63.
Figure 7-63 Select external trigger type in instrument setup window
4 Save the method by clicking File >> Method >> Save.
5 Download the method by clicking Control >> Download Method.
6 Enter Sample ID and enter Number of reps. See Figure 7-64.
NOTE: When multiple sample injections are defined in the Single Run
Acquisition window, it is required that each run be triggered by a
REMOTE_START contact closure by an external system. Once a run
is submitted, it will not start until the trigger is activated. Refer to section
5.2, The Remote Connector, on page 5-30.
Figure 7-64 Enter multiple samples into single run acquisition window