3000 Micro GC Operating Manual
8.2.4 Decommission a Standard GC Module
A Standard GC module must be decommissioned within the embedded web page
before it can be removed. 3000 Micro GC internally communicates to each
installed GC module using a unique module address (see Figure 8-3).
For a 1-2 channel or portable 3000 Micro GC, the module addresses are
1 and 2.
For a 3-4 channel 3000 Micro GC, the module addresses are 1, 2, 3 and 4.
Decommissioning a standard GC module will disable the GC module's address.
Figure 8-3 Default GC module serial communication addresses
NOTE: For Performance Enhanced modules, skip to section 8.2.5, Remove a GC
Module, on page 8-11.
NOTE: Only one Standard GC Module can be decommissioned at a time.
The procedure below describes the procedure for replacing the Standard GC
module 1 (Channel A) in a 2-channel 3000 Micro GC. The process is similar for
other configurations.
1 Open the web browser and type in the 3000 Micro GC IP Address.
The embedded webpage will display.
NOTE: At the factory, the instrument IP address was set to 10.1.1.101.
2 Click the Utilities tab.
3 Click Change Module Configuration. See Figure 8-4.
Address
1
1
2
234
Front
Front
2-channel 3000
Micro GC (top down view)
4-channel 3000
Micro GC (top down view)
Module
address
Corresponding
channel
1
2
3
4
A
B
C
D