3000 Micro GC Operating Manual
7.3 Configuring EZ IQ to Use a Valco (VICI) Valve
To configure an EZ IQ data system to use a Valco (VICI) valve:
1 Click Start >> All Programs >> Chromatography >> EZ IQ Config.
2 The EZ IQ Configuration window will display. Click Instrument
Configuration....See Figure 7-5.
Figure 7-5 EZ IQ configuration window
3 On the Instrument Configuration window, type the desired Instrument
name:, select Instrument type: 3000 Micro GC Enhanced Driver, and click
Configure. See Figure 7-6.
Figure 7-6 Instrument configuration window
For a Valco (VICI) Switching Valve, follow instructions in Configuring a Valco (VICI)
Switching Valve, see section 7.3.1 on page 7-5.
For a Valco (VICI) Sampling Valve, follow instructions in Configuring a Valco (VICI)
Sampling Valve, see section 7.3.2 on page 7-7.