4 - 11
PN 074-519-P1C
3000 Micro GC Operating Manual
4.2.4.5 Connecting the Carrier Gas to the 3000 Micro GC
1 Connect the carrier gases to the instrument via the 1/8 in. Swagelok fitting(s)
located on the back of the instrument. See section 4.2.3 on page 4-6.
WARNING
If a hazardous sample is used, the sample exhaust must
be vented to a fume hood or other designated area for
hazardous waste disposal.
CAUTION
Do not use leak detection fluids.
4.2.4.6 Using the Onboard Carrier Gas Cylinder (Portable 3000 Micro GC only)
For information on refilling the internal carrier gas cylinder, see section 8.5.3.1,
Refill Onboard Carrier Gas Cylinder, on page 8-43.
1 Remove the shipping caps.
2 Ensure the carrier gas knob on the front of the instrument is set to OFF.
See Figure 4-6.
Figure 4-6 Carrier gas knob
3 Connect the CARRIER OUT port to the CARRIER IN port with the carrier
jumper tube as show in Figure 4-7. Use CARRIER IN port 1 if 3000 Micro GC
has two ports available.