20    tirb45e1 (2019-02) CDGxxxDxx.cp 
The Low Trip Point is a fixed value: 
 
The value defined in Low Trip Point Limit is compared with the pressure value 
referenced by the Source Index parameter. 
 
 
 
 
Calculating the High Trip Point and Low Trip Point → chapter 4.5.2. 
 
  0x13  REAL  x  RW    Percentage High Trip 
 
 
0 
Disable 
 
 
0 
Disable 
 
 
High Trip Point Limit: High limit to trigger trip point condition if Input Value 
(I 0x900E, SI 0x01) is above this limit. 
 
 
Percentage of High Trip Source: Percentage of Piezo Sensor Value (I 0x6010, 
SI 0x11) in [%]. 
 
 
Low Trip Point Limit: Low limit to trigger trip point condition if Input Value (I 0x900E, 
SI 0x01) is below this limit. 
 
High Trip Hysteresis: Hysteresis value for High Trip Point. 
 
High Trip Point
Time t
in-
active
activein-
active
active
Low Trip Point
Measurement signal
(Pressure p)
Low Trip
Hysteresis
High Trip
Hysteresis
in-
active
4.5.3  Configuration Trip