CUSA® Clarity System Service and Maintenance Manual Page 39 of 117 0645403-2-EN
CAUTION
To avoid product damage, always use the torque base to hold the handpiece while using the torque wrench to tighten or loosen the tip. Avoid over
torqueing the tip.
6. Twist off the nosecone counter-clockwise until it unlocks.
7. Slide the nosecone off the tip.
8. Place the handpiece on the torque base as shown and make sure that the two flat metal sides fit snugly in the metal slot at the end of the
base.
9. Hold the handpiece in the base and place the torque wrench (color side towards handpiece) over the tip.