CUSA® Clarity System Service and Maintenance Manual Page 40 of 117 0645403-2-EN
10. Rotate the torque wrench counter-clockwise.
11. Unthread the tip from the handpiece connecting body by turning the tip counter-clockwise by hand.
12. Discard the items listed in Table 6-1 in compliance with hospital policy for contaminated waste.
Detaching the Console
1. Detach the aspiration tubing from the patient port on the canister.