CUSA® Clarity System Service and Maintenance Manual Page 46 of 117 0645403-2-EN
Package the handpiece and components for sterilization. Integra provides a sterilization tray for steam sterilization of the CUSA® Clarity handpiece.
This tray protects the handpiece during sterilization and during transfer to the sterile field.
Packaging the Handpiece and Components for Sterilization
The handpiece and components must be cleaned prior to sterilization, see “Cleaning the Handpiece, EUT, and Torque Base” in Section 6.3.1.
Sterilization Tray Layout
Handpiece
Flue and Stylet
Handpiece Cable Posts
Torque Base
Handpiece Connector
EUT Slots (3)
Nosecone
Lid (Not shown)
Most Common Sterilization Tray Configuration