Sterilizing the Handpiece and Components with Steam
Sterilizing the handpiece with steam depends on the following factors:
• Temperature
• Exposure time
• Population and resistance of resident bioburden
• Method of air removal from the autoclave
Use the validated steam sterilization cycle parameters outlined in these instructions. If you deviate from this recommended method of sterilizing,
it is your Health Care Facility’s responsibility to validate the deviation.
CAUTION
Do not use sterilization methods involving temperatures in excess of 134°C (273.2°F) as this may damage the handpiece.