D-0112389-I – 2022/04
Sera™ - Instructions for Use - EN Page 69
7 General technical specification
7.1 Sera
TM
instrument – technical specifications
Medical CE-
mark
The CE-mark in combination with MD symbol indicates that Interacoustics A/S
meets the requirements of the Medical Device Regulation (EU) 2017/745.
Approval of the quality system is made by TÜV no. 0123.
Standards Safety: IEC 60601-1:2012, Internally powered, Type B and BF
EMC: IEC 60601-1-2:2014
Calibration: ISO 389-2:1994
Cradle
Safety:
Power
Mains voltages and
frequencies:
IEC 60601-1:2012, Class II
UES12LCP-050160SPA Item number 8515473
100 – 240 V~, 50/60 Hz, 400 mA
Operation
environment
Temperature:
Relative Humidity:
Ambient Pressure:
Boot-up time:
5 – 40°C, + 41°F... + 104°F
15 – 93% (non-condensing)
98 kPa – 104 kPa
< 5 sec
environment
Transport
Temperature:
Storage and Transport
0°C – 50°C, - 4°F... + 122°F
-25 – 70°C, - 13°F... + 158°F
Max 93% (non-condensing)
2000 m / 6561 f t above sea level
IP02
IP20
IP marking is an ingress protection marking. The marking specif ies the protection
provided against ingress of particle matter and liquids. This instrument has
different IP marking with the follow impact:
IP02: To protect the instrument against rain and water always
during transport.
IP20: This marking can be f ound on the instrument parts meaning that the parts
are not protected against water
NOTE: The charger, power supply and cradle are not to be used in home