Sera™ - Instructions for Use - EN Page 49
4.8.4.1 Probe stability % (DPOAE only)
When the measurement is finished, the Probe stability % provides an indication of how stable the probe
placement was throughout the entire test. A low probe stability % indicates that the probe was moving
during testing.
4.8.4.2 Stimulus stability % (TEOAE only)
When the measurement is finished, the Stimulus stability % provides an indication of how stable the
stimulus level was throughout the measurement. A low stability value usually indicates that the probe was
moving during testing.
4.8.4.3 Artifact %
When the measurement is finished, the total amount of artif act measured during the test will be shown.
4.8.4.4 Time
When the measurement is finished, the total test time will be presented here.