D-0112389-I – 2022/04
Sera™ - Instructions for Use - EN Page 71
Standards)
OWA Probe
TM
Rejection level (Peak, Min RMS, Max RMS) & Clipping
(Saturation)
Transducer, test result bars, result symbols
(pass/refer/incomplete), test time, artifact, electrode
Measurement
(default protocol):
Please refer to the Sera
TM
Additional Information
for further inf ormation about
available test
, test protocol settings and s
ensitivity and
specificity.
SnapPROBE
TM
auto detection, auto calibrated
Replaceable probe tip
OWA Probe auto detection, auto calibrated
24 bit, 5.38 Hz resolution
Artifact (noise)
rejection system:
An RMS average measurement in the DP-bin f requency
area (26 bins at frequencies < 2500 Hz & 60 bins ≥
Transducer, test result bars, result symbols
(pass/refer/incomplete), probe stability, test time,
Measurement:
(default protocol):
Please refer to the Sera
TM
Additional Information
for further inf ormation about available test
protocols, test protocol settings and sensitivity and
specificity.
The Sera
TM
with DPOAE uses an improved method of stimuli level control, which more accurately delivers
the specif ied intensity in the f ull range of ear canals, f rom infants to adults. The applicability of the IEC
60645-6 standard is currently limited to adult ears. Therefore, to better serve a market with a product that
provides more accurate stimulus levels to a wide range of ear canal volumes (specif ically infants), we have
elected to utilize a more comprehensive calibration procedure for DPOAEs that is outside the scope of
IEC 60645-6 for some protocols.
All protocols that include IEC in the name use the specified IEC calibration method.