4210 Flow Meter
Section 4 Optional Equipment
4-31
Dissolved Oxygen, mg/L
Sensor Type Calculated from % air saturation, tem-
perature, and salinity
Range 0 to 20 mg./L
Accuracy ± 0.2 mg./L
Resolution 0.01 mg./L
Conductivity*
Sensor Type Four-Electrode Cell
Range 0 to 100 mS/cm
Accuracy ± (1% of reading + 0.001 mS/cm)
Resolution 0.01 mS/cm or 1 µS/cm
*Specific conductance (conductivity corrected to 25° C),
resistivity, and total dissolved solids measurements are
also provided. These values are automatically calcu-
lated from conductivity according to algorithms found
in Standard Methods for the Examination of Water and
Wastewater (ed. 1989).
Salinity
Sensor Type Calculated from conductivity and tem-
perature
Range 0 to 70 ppt
Accuracy ± 0.2 ppt
Resolution 0.1 ppt
pH
Sensor Type Glass Combination Electrode
Range 2 to 14 pH Units
Accuracy ± 0.2 Units
Resolution 0.01 Units
Table 4-6 YSI Sensor Technical Specifications