4210 Flow Meter
Section 2 Programming
2-46
outside of “normal.” If the enabling condition returns to
“normal,” the sampler enable will turn off until the next time the
condition goes outside of “normal.
In the latching mode, the sampler will be enabled the first time
the condition goes outside of normal and the sampler will remain
enabled, regardless of any subsequent changes to the enabling
condition. If this is the case, the following menu will allow you to
reset the sampler enable feature. Note that this menu will not
appear unless the condition necessary to enable the sampler has
been met and the sampler is currently enabled.
Or.
Select YES to reset the sampler enable feature; select NO to
leave the sampler enabled.
This selection allows you to turn the flow meter's internal printer
on or off when the sampler is enabled from the flow meter. This
allows you to conserve battery power and only print a chart when
the sampler is enabled. This feature is useful for monitoring
storm water runoff.
2.12 Step 7 - Alarm Dialout
Mode
This step lets you signal an alarm to a remote location from the
flow meter. The conditions that can cause an alarm are the same
as described previously for sampler enabling.
Note
You must have the optional internal modem installed and con-
nected to a telephone network to make use of this feature. This
menu will not even appear if you do not have a modem. The
flow meter will automatically advance to the next step.
More information about the modem can be found in Section 4.2.
If you have the modem installed and select step 7, the following
will appear:
If you select DISABLE, this option will be deactivated until you
change the selection later. The program will advance to the next
step. If you select STORM, the flow meter will request definitions
similar to those for STORM in sampler enable. The following will
appear:
WHEN ENABLE CONDITION IS NO LONGER MET
• DISABLE SAMPLER • • KEEP ENABLED •
ENABLE CURRENTLY LATCHED, RESET
• NO • • YES •
PRINTER ON/OFF WITH ENABLE
• YES • • NO •
ALARM DIAL OUT
• DISABLE • • CONDITIONAL • • STORM • • FLOWLINK •
LEVEL
GREATER THAN X.XXX FT (or other units of measure)