4210 Flow Meter
Section 2 Programming
2-40
Figure 2-2 YSI 600 Sonde Calibration Flow Chart
Prog ram St ep #3, “PARAMETER TO ADJUST"
Parame ter To Adjust
None - Level-YSI 600 - Ph - D.o.
YSI 600 Para
met e r To C alibr a te
None - Ph - Conductivity - D.o.
YS I 600 Ph Calibration
pH4 & pH7 - pH7 & pH 10 - pH4, pH7, & pH10
Ri n s e
probe and place in ph 4.0 solution
Press Enter when stable: X.XX pH
Calib rating .
..
Please Wa it...
Error/warning Override
Yes - No
Rinse probe and place in ph 7.0 solution
Press Enter when ready: X. XX pH
Calib rating ...
Please Wa it...
Er ror/warn ing Ove rrid e
Yes - No
Rinse Probe And Place In Ph 7.0 Solution
Press En ter when stable: X. XX pH
Calib rating ...
Ple ase Wa it...
Error/warning Override
Yes - No
Rinse probe and place in ph 10.0 solution
Press Enter when stable: X.XX pH
Err or/warning Override
Yes - No
Rinse probe and place in ph 4.0 solution
Press Enter when stable: X.XX pH
Er ror/warn ing overrid e
Yes - No
Rinse probe and place in ph 7.0 solution
Press En ter when stable: X. XX pH
Error/warning override
Yes - No
Ri nse probe and place in ph 10.0 solution
Press Enter when stable: X.XX pH
Er ro r/wa rning ove rride
Yes - No
No
Yes
No
Yes
Good Ca libratio n
E rro r/
Wa rn i n g
Erro r /
Wa r n i n g
Er ro r/
Wa rn i n g
Erro r/
Wa
rn in g
Erro r/
Wa r n i n g
Erro r /
Wa r n i n g
E rro r/
Wa rn i n g
Good Cali brat ion
Good Cali bration
Good Ca librat ion
Good Ca librat ion
Good Cali brat ion
Good Ca librat ion
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Calib rating ...
Please Wa it...
Calib rating .
..
Please
wait...
Calibra ting...
Please wait...
Ca libratin g ...
Ple ase wait...
YS I 600 Parameter To Calibrate
Non e - Ph - Cond uctivity - D.o.
YS I 600 D.O. Calibration
D.O. Standard - Abs. Barometric Pressure - Altitude
Place Probe in X.XX Mg / l solu t io n
Press Enter when ready:
Er
ror/warning Override
Yes - No
Units for altitude entry
Ft . - M
Er ror/warning override
Yes - No
Enter conductivit y ca lib ratio n un its
mS/cm - ppt
Error/warning Override
Yes - No
No
Yes
No
Yes
Good Cali brat ion
Erro r/
Wa r n i n g
Er ro r/
Wa rn i n g
Erro r/
Wa r n i n g
Good Cali brat ion
Good Calib ration
No
Yes
D.O. Stan d ard
X.XX Mg / l
Ab s . Baro m e t ric Pr e ss u re
X.XX mmHg
Place prob e in calib ration cu p
Press En ter when ready:
Error/warning Override
Yes - No
No
YesGood Calib ration
Erro r/
Wa r n i n g
Place p robe in calibratio n cup
Pre ss En te r when ready:
En ter co n d u ct ivity standard
X XX Un it s
Place probe in XXX units solution
Press Enter when ready
Calib rating ...
Press Enter when stable X.XX Mg / l
Calibra ting...
Press Enter when stable X. XX Mg / l
Calibratin g ...
Press Enter when stable X.XX mS /cm
Ca li
brating ...
Press En ter when stable X.XX Mg / l