2 Instrument description
The time period for the counter/integrator summation will not be freshly start-
ed. The previous counter/integrator values will also not be saved.
In this way, you can restart the recording of individual counters/integrators, for
example, after the test run of an installation; the values of the test run, which
are not required, can thus be eliminated.
You can select a password other than that for accessing the configuration. The
password can be set under Configuration
Instrument data
Code
No.(Password)
Counter/Int.reset.
Reset via
the “Disk
manager” menu
If the function Update diskette incl. counters is executed in the “Disk manager”
menu, the counts will also be stored and reset.
v See “Disk manager” on page 38.
2.5.3 Behavior on instrument reconfiguration
When the instrument is reconfigured, the current counter/integrator reporting
periods remain unaffected. The counter/integrator values will not be reset to 0
and the reporting period will not be restarted.
H
If you wish to save the previous counter/integrator values, you have to
execute the function “Update diskette incl. counters” in the Disk manager
menu before resetting.
H
The values can be deliberately reset via the “Parameters” menu.