4 Configuration parameters
60
4.2.7 Configuration - Outputs (option)
Start time
timed operation
Configuration
➔ Meas. storage
➔ Timed operation
➔ Start time
any time Off when
start time =
end time
End time
timed operation
Configuration
➔ Meas. storage
➔ Timed operation
➔ End time
any time
Stored value
timed operation
Configuration
➔ Meas. storage
➔ Timed operation
➔ Stored value
Average val.,
Instant val.,
Minimum,
Maximum,
Peak value
v Chapter 2.7 “Operating
modes”
Chapter 2.8 “Data
storage”
Storage rate
timed operation
Configuration
➔ Meas. storage
➔ Timed operation
➔ Storage rate
1—5 — 32767s v Chapter 2.7 “Operating
modes”
Chapter 2.8 “Data
storage”
Parameter Value/selection Description
Action
outputs
Configuration
➔ Outputs
➔ Relay K1
➔ Action
Off,
n.o. (make),
n.c. (break)
Operating signal
outputs
Configuration
➔Outputs
➔ Relay K1
➔ Operating signal
Off,
Logic inp1 — 4,
Logic channel 1 — 6,
Low alarm 1 — 6,
Low comb. alarm,
High alarm 1 — 6,
High comb. al.,
Counter/I al. 1 — 6,
C/I comb. al.,
Comb. alarm,
Memory al.,
Error,
Modbus-Flag
The configured signal is
output to the relay.
Action
outputs
Configuration
➔Outputs
➔ Relay K2 — K3
➔ Action
Off,
n.o. (make),
n.c. (break)
If n.o. (break) or n.o. (make)
is selected, the parameter
Outputs is automatically
set to Low alarm 1 or Low
alarm 2.