3 Operation and visualization
42
Instrument info
- This menu also includes the version number (e.g. 133.01.01), i.e. the
version number of the instrument software. It is important because some
functions are only available from a certain version number onwards.
- If the “Digital-I/O“ is available (Yes), then the recorder also contains an
RS232/RS485 interface.
Error The following errors are possible:
Error Description
none Instrument o.k.
Data lost A discharge of the battery/storage capacitor
occured during the last lengthy supply
interruption.
The clock was set to 01.01.97 00:00:00.
Remedy:
Reset the time (v Chapter 4.2.2 “Configuration -
Instrument data”).
Battery empty This message appears on instruments with lithium
battery when the time was reset after a data loss.
Please return the instrument to the supplier for a
change of battery.
H
Data can be lost after separating the instrument from the supply
voltage after more than 10 years on instruments with a lithium bat-
tery, and after more than 2 weeks (15 — 25°C ambient tempera-
ture) on instruments with a storage capacitor.