51
4 Configuration parameters
Memory alarm
(diskette
reserve)
Configuration
➔ Instrument data
➔ Memory alarm
1—10 — 100% The signal is activated
when the residual capacity
of the diskette, or of the
internal memory, has
fallen to this value.
The parameter Memory
readout can be used to
determine whether the
alarm was initiated by the
diskette or the memory.
v Chapter 2.4 “Event tra-
ces”
Code number
of configuration
Configuration
➔ Instrument data
➔ Code No. (password)
➔ Configuration
0000 — 9200 — 9999 Code number for
configuration level;
0000 = off
The data saved on dis-
kette and in the FLASH
memory are deleted
when the configuration is
changed.
Code number
of disk manager
(Disk Code No.)
Configuration
➔ Instrument data
➔ Code No. (password)
➔ Disk manager
0000 — 9200 — 9999 Code number for
functions in the
“File manager” menu;
0000 = off
Code number
Counter/Int. reset
Configuration
➔ Instrument data
➔ Code No. (password)
➔ Counter/Int. reset
0000 — 9200 — 9999 Code number to delete the
individual counts;
0000 = off
Code number
RS232/RS485
Configuration
➔ Instrument data
➔ Code No. (password)
➔ RS232/RS485
0000 — 9999 Code number to protect
from unauthorized access
to data via the serial
interface;
0000 = off
Factory setting Configuration
➔ Instrument data
➔ Enter defaults
No, Yes Yes = enter factory default
setting (when entered, the
parameter returns
automatically to No)
Enable options Configuration
➔ Instrument data
➔ Enable options
The parameter is available
for enabling additional
functions, such as the
math/logic module or
counter/integrator. The
parameter is not available
if all recorder functions
have been enabled.
Enable options
Code No.
determined
Configuration
➔ Instrument data
➔ Enable options
➔ Code No.
determined
(display of Code No.) The manufacturer has to
be informed about the
value displayed here and
will issue the enabling
code.