EasyManua.ls Logo

Kamstrup MULTICAL 402 - 4.2 Mounting of flow sensor

Kamstrup MULTICAL 402
113 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MULTICAL® 402
12
Kamstrup A/S · Technical Description · 5512-742_M1_GB_01.2016
2.5 Accuracy
Heat meter components MPE according to EN 1434-1
MULTICAL
Flow sensor
402, typical accuracy
Ef = ± (2 + 0.02 q
p
Ef= ± (1 + 0.01 qp/q) % /q), but not exceeding ±5 %
Calculator
Ec= ± (0.5 + Θ
min
Ec= ± (0.15 + 2/Θ) % /∆Θ) %
Sensor pair
Et= ± (0.5 + 3 ∆Θ
min
Et= ± (0.4 + 4/Θ) % /∆Θ) %
Diagram 1: Total typical accuracy of MULTICAL
®
402 compared to EN 1434-1
-6,0
-4,0
-2,0
0,0
2,0
4,0
6,0
0,01 0,10 1,00 10,00
Tol [%]
q [m³/h]
MULTICAL
®
402 q
p
1,5 m³/h @
∆Θ
30K
Ec+Et+Ef
(EN)