10 Technical data
kV: +- 5kV
mA: +- 1mA / +-20%
time: +- 10%
DAP: +- 50%
Image storing and retrieving:
File formats PNG (16-bit), JPG (12-bit)
File compression PNG (lossless), JPG (100%-60% quality)
Panoramic file size 2-4 MB
Cephalometric file size 3-5 MB
3D file size 12-400 MB (DICOM)
Patient database Standalone workstation
Server on local area network (LAN)
Panoramic patient positioning
Operation Left or right side of unit. Motorised carriage
movement
Positioning aids Chin rest, bite block, 3-point headrest Curved
mirror, 3 positioning laser lights, Occlusion
correction buttons
Cephalostat patient positioning
Operation Arm mounts on left or right side of the unit
Interlocked pan/ceph sensor. Motorised carriage
buttons at cephalostat head assembly.
Positioning aids Ear rods, Nasion support with vertical mm scale,
Frankfort horizontal plane laser light, Contact plate
(Carpus program).
3D imaging patient positioning
Operation Left or right side of unit. Motorised carriage
movement
Positioning aids Chin rest, chin support, 3-point headrest, Curved
mirror, 3 positioning laser lights
Cephalostat scanning
Scanning method Horizontal scan, synchronized sensor and
secondary slot motion
Scanning time 10 - 20 s.
ORTHOPANTOMOGRAPH