4 Imaging programs
Exposure settings for cephalometric lateral program
100
VAC
90 kV/ 8 mA/ 10
s/ 30 mGycm
2
90 kV/ 8 mA/ 13
s/ 37 mGycm
2
90 kV/ 8 mA/ 16
s/ 48 mGycm
2
90 kV/ 8 mA/ 20
s/ 60 mGycm
2
120
VAC
85 kV/ 10 mA/ 10
s/ 34 mGycm
2
85 kV/ 10 mA/ 13
s/ 42 mGycm
2
90 kV/ 10 mA/ 16
s/ 60 mGycm
2
90 kV/ 10 mA/ 20
s/ 75 mGycm
2
230
VAC
85 kV/ 10 mA/ 13
s/ 42 mGycm
2
90 kV/ 10 mA/ 13
s/ 47 mGycm
2
90 kV/ 12.5 mA/
16 s/ 76 mGycm
2
90 kV/ 12.5 mA/
20 s/ 95 mGycm
2
Cephalo posterior-anterior (PA) projection
Lateral Cephalostat uses a full sensor image height.
Reverse towne projection Waters view
32 ORTHOPANTOMOGRAPH