4 Imaging programs
Exposure settings for cephalometric PA program
100 VAC 90 kV/ 8 mA/ 10
s/ 30 mGycm
2
90 kV/ 8 mA/ 13
s/ 37 mGycm
2
90 kV/ 8 mA/ 16
s/ 48 mGycm
2
90 kV/ 8 mA/ 20
s/ 60 mGycm
2
120 VAC 85 kV/ 10 mA/ 10
s/ 34 mGycm
2
85 kV/ 10 mA/ 13
s/ 42 mGycm
2
90 kV/ 10 mA/ 16
s/ 60 mGycm
2
90 kV/ 10 mA/ 20
s/ 75 mGycm
2
230 VAC 85 kV/ 10 mA/ 13
s/ 42 mGycm
2
90 kV/ 10 mA/ 13
s/ 47 mGycm
2
90 kV/ 12.5 mA/
16 s/ 76 mGycm
2
90 kV/ 12.5 mA/
20 s/ 95 mGycm
2
Carpus view (optional)
Carpus holder is a separate option.
Not available in USA and Canada.
Exposure settings for Carpus view program
100 VAC
120 VAC
230 VAC
66 kV/ 3,2 mA/ 8
s/ 5 mGycm
2
70 kV/ 3,2 mA/ 8
s/ 5 mGycm
2
73 kV/ 3,2 mA/ 8
s/ 6 mGycm
2
73 kV/ 6,3 mA/ 8
s/ 12 mGycm
2
ORTHOPANTOMOGRAPH
OP 3D Pro 33