SourceMeter
®
Line Specifications
RESISTANCE MEASUREMENT ACCURACY (Local or Remote Sense)
DEFAULT DEFAULT NORMAL ENHANCED
TEST TEST CURRENT ACCURACY (23°C ±5°C) ACCURACY (23°C ±5°C)
5
DEFAULT CURRENT 2420, 2425, 1 YEAR,±(% rdg.+ ohms) 1 YEAR, ±(% rdg. + ohms)
RANGE RESOLUTION 2400, 2410 2430, 2440 2400 2410 2420, 2425, 2430, 2440 2400
<0.20000 Ω
4
– – – Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
2.00000 Ω
4
10 µΩ – 1 A Source I
ACC
+ MeasV
ACC
Source I
ACC
+ Meas.V
ACC
0.17% +0.0003 Ω Source I
ACC
+ Meas.V
ACC
20.0000 Ω 100 µΩ 100 mA 100 mA 0.10% + 0.003 Ω 0.11% + 0.006 Ω 0.10% + 0.003 Ω 0.07% + 0.001 Ω
200.000 Ω 1mΩ 10 mA 10 mA 0.08% + 0.03 Ω 0.09% + 0.1 Ω 0.08% + 0.03 Ω 0.05% + 0.01 Ω
2.00000 kΩ 10 mΩ 1 mA 1 mA 0.07% + 0.3 Ω 0.08% + 0.6 Ω 0.07% + 0.3 Ω 0.05% + 0.1 Ω
20.0000 kΩ 100 mΩ 100 µA 100 µA 0.06% + 3 Ω 0.07% + 6 Ω 0.06% + 3 Ω 0.04% + 1 Ω
200.000 kΩ 1 Ω 10 µA 10 µA 0.07% + 30 Ω 0.07% + 60 Ω 0.07% + 30 Ω 0.05% + 10 Ω
2.00000 MΩ
8
10 Ω 1 µA 1 µA 0.11% + 300 Ω 0.12% + 600 Ω 0.11% + 300 Ω 0.05% + 100 Ω
20.0000 MΩ
9
100 Ω 1 µA 1 µA 0.11% + 1 kΩ 0.12% + 2.4 kΩ 0.11% + 1kΩ 0.05% + 500 Ω
200.000 MΩ
4
1kΩ 100 nA – 0.66% + 10 kΩ 0.66% + 24 kΩ Source I
ACC
+ Meas.V
ACC
0.35% + 5 kΩ
>200.000 MΩ
4
– – – Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
Source I
ACC
+ Meas.V
ACC
VOLTAGE MEASUREMENT ACCURACY (Local or Remote Sense)
ACCURACY (1 Year)
DEFAULT INPUT 23°C ±5°C
MODEL RANGE RESOLUTION RESISTANCE ±(% rdg. + volts)
ALL: 200.000 mV 1 µV >10 GΩ 0.012% + 300 µV
2.00000 V 10 µV >10 GΩ 0.012% + 300 µV
2400: 20.0000 V 100 µV >10 GΩ 0.015% + 1.5 mV
200.000 V 1 mV >10 GΩ 0.015% + 10 mV
2410: 20.0000 V 100 µV >10 GΩ 0.015% + 1 mV
1000.00 V 10 mV >10 GΩ 0.015% + 50 mV
2420, 2425, 2430: 20.0000 V 100 µV >10 GΩ 0.015% + 1 mV
2420: 60.0000 V 1 mV >10 GΩ 0.015% + 3 mV
2430: 100.000 V 1 mV >10 GΩ 0.015% + 5 mV
2440: 10.0000 V 100 µV >10 GΩ 0.015% + 750 µV
40.0000 V 1 mV >10 GΩ 0.015% + 3 mV
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15 × accuracy specification)/°C.
CURRENT MEASUREMENT ACCURACY (Local or Remote Sense)
ACCURACY (1 Year)
DEFAULT VOLTAGE 23°C ±5°C
MODEL RANGE RESOLUTION BURDEN
3
±(% rdg. + amps)
2400, 2410: 1.00000 µA 10 pA < 1mV 0.029% + 300 pA
All: 10.0000 µA 100 pA < 1mV 0.027% + 700 pA
100.000 µA 1 nA < 1mV 0.025% + 6 nA
1.00000 mA 10 nA < 1mV 0.027% + 60 nA
2400, 2420, 2425, 2430, 2440: 10.0000 mA 100 nA < 1mV 0.035% + 600 nA
2410: 20.0000 mA 100 nA < 1mV 0.035% + 1.2 µA
All: 100.000 mA 1 µA < 1mV 0.055% + 6 µA
2400, 2410: 1.00000 A 10 µA < 1mV 0.22% + 570 µA
2420, 2425, 2430, 2440: 1.00000 A 10 µA < 1mV 0.060% + 570 µA
2420, 2425: 3.00000 A 10 µA < 1mV 0.052% + 1.71 mA
2430: 3.00000 A 10 µA < 1mV 0.052% + 1.71 mA
10.0000 A
6
2440: 5.00000 A 10 µA < 1mV 0.10% + 3.42 mA
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.10 × accuracy specification)/°C.
MEASURE SPECIFICATIONS
1, 2,7
1
Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200mV, 1A, 10A
ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200mV, 1A, 10A
ranges, add 0.5%.
2
Accuracies apply to 2- or 4-wire mode when properly zeroed.
3
4-wire mode.
4
Manual ohms only – except 2420, 2425, 2430, 2440 for 2Ω range and 2410 or 2400 for 200MΩ range.
5
Source readback enabled, offset compensation ON. Also available on 2410, 2420, 2425, 2430, and 2440
with similar accuracy enhancement.
6
10A range only in pulse mode.
7
In pulse mode, limited to 0.1 PLC measurement.
8
Except 2440; default test current is 5µA.
9
Except 2440; default test current is 0.5µA.
TEMPERATURE COEFFICIENT (0°–18°C & 28°–50°C): ±(0.15 × accuracy
specification)/°C.
SOURCE I MODE, MANUAL OHMS: Total uncertainty = I source accuracy
+ V measure accuracy (4-wire remote sense).
SOURCE V MODE, MANUAL OHMS: Total uncertainty = V source accuracy
+ I measure accuracy (4-wire remote sense).
6-WIRE OHMS MODE: Available using active ohms guard and guard
sense. Max. Guard Output Current: 50mA (except 1A range). Accuracy
is load dependent. Refer to White Paper no. 2033 for calculation
formula.
GUARD OUTPUT IMPEDANCE: <0.1Ω in ohms mode.
HW Rev. B
4/30/02
CONTACT CHECK
SPECIFICATIONS
SPEED: 350µs for verification and notification.
CONTACT CHECK: 2Ω 15Ω 50Ω
No contact
check failure <1.00Ω <13.5Ω <47.5Ω
Always contact
check failure >3.00Ω >16.5Ω >52.5Ω
2400, 2400-C,2410, 2410-C,
2420, 2420-C,2425, 2425-C,
2430, 2430-C,2440, 2440-C