AEJ-C/AES-C/PLJ-C-BA-e-1621 131
Use the navigation keys to enter the test volume <V1> (such as 100 µl); the
active digit will be flashing.
This completes the entry for pipettes with fixed volume. For pipettes with
variable volume select volume V2 and V3 as described below.
Confirm your entry by pressing the PRINT-key and the display for entering the
second test volume <V2> will be shown.
Use the navigation keys to enter the test volume <V2> (such as 500 µl), the
active digit will be flashing.
Confirm your entry by pressing the PRINT-key and the display for entering the
third test volume <V3> will be shown.
Use the navigation keys to enter the test volume <V3> (such as 1000 µl), the
active digit will be flashing.
Confirm your entry by pressing the PRINT-key and the display for performing
pipette calibration will be shown.
Follow the instructions in the informational line of text.