4 Features and Functions
106 Keysight 34970A/34972A User’s Guide
If a scan includes a read of the totalizer, multifunction module, the count is
reset each time it is read during the scan only when the totalizer reset mode is
enabled (TOTalize:TYPE RRESet command or the Advanced menu for the
totalizer).
If you install a module while a scan is running, the instrument will cycle power
and resume scanning. If you remove a module while a scan is running, the
instrument will cycle power, and it will not resume scanning when the reboot
completes. If you are logging to a USB drive, the instrument will not log any
scans that occur between the removal of the module and the reboot of the
instrument.
You can use either the internal DMM or an external DMM to make
measurements of your configured channels. However, the instrument allows
only one scan list at a time; you cannot scan some channels using the internal
DMM and others using an external DMM. Readings are stored in 34970A/
34972A memory only when the internal DMM is used.
If the internal DMM is installed and enabled, the instrument will automatically
use it for scanning. For externally-controlled scans, you must either remove
the internal DMM from the 34970A/34972A or disable it (see “Internal DMM
Disable” on page 157).