Features and Functions 4
Keysight 34970A/34972A User’s Guide 201
Calibration message
The instrument allows you to store one message in calibration memory in the
mainframe. For example, you can store such information as the date when the last
calibration was performed, the date when the next calibration is due, the
instrument’s serial number, or even the name and phone number of the person to
contact for a new calibration.
You can record a calibration message only from the remote interface and only
when the instrument is unsecured. You can read the message from either the
front-panel or over the remote interface. You can read the calibration message
whether the instrument is secured or unsecured.
The calibration message may contain up to 40 characters. From the front
panel, you can view 13 characters of the message at a time. Press to scroll
through the text of the message. Press again to increase the scrolling
speed.
Storing a calibration message will overwrite any message previously stored in
memory.
The calibration message is stored in non-volatile memory in the mainframe,
and does not change when power has been off, after a Factory Reset (*RST
command), or after an Instrument Preset (SYSTem:PRESet command).
Front-Panel Operation:
CAL MESSAGE
Remote Interface Operation: To store the calibration message, send the
following command.
CAL:STRING ’CAL: 06-01-98’

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight 34972A and is the answer not in the manual?

Keysight 34972A Specifications

General IconGeneral
Model34972A
CategoryData Acquisition/Switch Unit
ManufacturerKeysight Technologies
Scan RateUp to 250 channels/sec
Voltage Measurement Range100 mV to 300 V
Resistance Measurement Range100 Ω to 100 MΩ
Operating Temperature0 °C to 55 °C
Storage Temperature-40 °C to 70 °C
Switching ModulesSupports plug-in modules
Current Measurement Range1 µA to 1 A
Temperature MeasurementThermocouples, RTDs, and Thermistors
InterfaceUSB, LAN
Memory50, 000 readings
Weight3.6 kg