5 Error Messages
248 Keysight 34970A/34972A User’s Guide
Calibration Errors
The following errors indicate failures that may occur during a calibration. Refer to
the 34970A/34972A Service Guide for more information.
701 Cal: security disabled by jumper
The calibration security feature has been disabled with a jumper inside the
instrument. When applicable, this error will occur at power-on to alert you that
the instrument is unsecured.
702 Cal: secured
The instrument is secured against calibration.
703 Cal: invalid secure code
You have entered an invalid calibration security code. You must use the same
security code to unsecure the instrument that was used to secure it, and vice
versa. The security code may contain up to 12 alphanumeric characters. The first
character must be a letter, but the remaining characters can be letters,
numbers, or an underscore ( _ ). You do not have to use all 12 characters but the
first character must always be a letter. The security code is set to “KT034970” or
“KT034972” when the instrument is shipped from the factory.
704 Cal: secure code too long
The security code may contain up to 12 alphanumeric characters. A security
code was received which contained more than 12 characters.
705 Cal: aborted
A calibration in progress is aborted when you turn off the instrument or send a
bus Device Clear.
706 Cal: value out of range
The specified calibration value (CALibration:VALue) is not valid for the present
measurement function and range.
707 Cal: signal measurement out of range
The specified calibration value (CALibration:VALue) does not match the signal
applied to the instrument.