EasyManua.ls Logo

Keysight 34972A - Thermistor Measurements

Keysight 34972A
353 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Features and Functions 4
Keysight 34970A/34972A User’s Guide 141
Thermistor measurements
To connect a thermistor to the module’s screw terminals, see page 39
The instrument supports 2.2 kΩ (44004), 5 kΩ (44007), and 10 kΩ (44006)
thermistors.
Front-Panel Operation: To select the thermistor function for the active
channel, choose the following items.
TEMPERATURE , THERMISTOR
To select the thermistor type for the active channel, choose from the following
items.
TYPE 2.2 KOHM , TYPE 5 KOHM , TYPE 10 KOHM
Remote Interface Operation: You can use the MEASure? or CONFigure
command to select the probe type and thermistor type. For example, the
following statement configures channel 301 for measurements of a 5 kΩ
thermistor:
CONF:TEMP THER,5000,(@301)
You can also use the SENSe command to select the probe type and thermistor
type. For example, the following statement configures channel 103 for
measurements of a 10 kΩ thermistor:
SENS:TEMP:TRAN:THERM:TYPE 10000,(@103)

Table of Contents