Thermistor measurements
To connect a thermistor to the module’s screw terminals, see page 39
– The instrument supports 2.2 kΩ (44004), 5 kΩ (44007), and 10 kΩ (44006)
thermistors.
– Front-Panel Operation: To select the thermistor function for the active
channel, choose the following items.
TEMPERATURE , THERMISTOR
To select the thermistor type for the active channel, choose from the following
items.
TYPE 2.2 KOHM , TYPE 5 KOHM , TYPE 10 KOHM
– Remote Interface Operation: You can use the MEASure? or CONFigure
command to select the probe type and thermistor type. For example, the
following statement configures channel 301 for measurements of a 5 kΩ
thermistor:
CONF:TEMP THER,5000,(@301)
You can also use the SENSe command to select the probe type and thermistor
type. For example, the following statement configures channel 103 for
measurements of a 10 kΩ thermistor:
SENS:TEMP:TRAN:THERM:TYPE 10000,(@103)