4 Features and Functions
140 Keysight 34970A/34972A User’s Guide
You can also use the SENSe command to select the probe type, RTD type, and 
nominal resistance. For example, the following statement configures channel 
103 for 4-wire measurements of an RTD with 
α = 0.00391 (channel 103 is automatically paired with channel 113 for the 
4-wire measurement).
SENS:TEMP:TRAN:FRTD:TYPE 91,(@103)
 
The following statement sets the nominal resistance (R
0
) to 1000Ω on channel 
103.
SENS:TEMP:TRAN:FRTD:RES 1000,(@103)