4 Features and Functions
190 Keysight 34970A/34972A User’s Guide
Remote interface selection
The 34970A is shipped with both an GPIB (IEEE-488) interface and an RS-232
interface. Only one interface can be enabled at a time. The GPIB interface is
selected when the instrument is shipped from the factory.
The interface selection is stored in non-volatile memory, and does not change
when power has been off, after a Factory Reset (*RST command), or after an
Instrument Preset (SYSTem:PRESet command).
If you select the GPIB interface, you must select a unique address for the
instrument. The GPIB address is displayed on the front panel when you turn on
the instrument.
If you select the RS-232 interface, you must also set the baud rate, parity, and
flow control mode for the instrument. “RS-232” is displayed on the front panel
when you turn on the instrument.
Front-Panel Operation:
GPIB / 488 , RS-232
Remote Interface Operation:
SYSTem:INTerface {GPIB|RS232}
Baud rate selection (RS-232)
You can select one of eight baud rates for RS-232 operation. The rate is set to
57,600 baud when the instrument is shipped from the factory.
You can set the baud rate from the front panel only.
Select one of the following: 1200, 2400, 4800, 9600, 19200, 38400, 57600
(factory setting), or 115200 baud.
The baud rate selection is stored in non-volatile memory, and does not change
when power has been off, after a Factory Reset (*RST command), or after an
Instrument Preset (SYSTem:PRESet command).
Front-Panel Operation:
19200 BAUD