20 AWG Typical
6 mm
2 Connect wiring to the screw terminals.
.
Slot Channel
Channel Number:
1 Remove the module cover.
3 Route wiring through strain relief.
5 Install the module into mainframe.
Cable Tie Wrap
(optional)
4 Replace the module cover.
Wiring Hints...
– For detailed information on each module,
refer to the section starting on page 209.
– To reduce wear on the internal DMM relays,
wire like functions on adjacent channels.
– For information on grounding and shielding,
see page 270.
– The diagrams on page 38 show how to
connect wiring to a multiplexer module for
each measurement function.