Features and Functions 4
Keysight 34970A/34972A User’s Guide 175
Real-time system clock
During a scan, the instrument stores all readings and alarms with the current time
and date. The instrument stores the time and date information in non-volatile
memory.
When shipped from the factory, the instrument is set to the current time and
date (U.S. Mountain Time).
Front-Panel Operation:
TIME 03:45 PM
J
UN 01 1997
Remote Interface Operation: Use the following commands to set the time and
date.
Internal DMM disable
You can scan through the configured channels using either the internal DMM or
an external instrument. For externally-controlled scans, you must either remove
the internal DMM from the instrument or
For information on controlling a scan with an external instrument, refer to
“Scanning with External Instruments” on page 125.
When shipped from the factory, the internal DMM is enabled. When you
change the state of the internal DMM, the instrument issues a Factory Reset
(*RST command).
–A Factory Reset (*RST command) or Instrument Preset (SYSTem: PRESet
command) does not affect the internal DMM configuration.
Front-Panel Operation:
DMM ENABLED , DMM DISABLED
Remote Interface Operation:
INSTrument:DMM {OFF|ON}
SYST:TIME 15,45,00 Set time to 3:45 PM
SYST:DATE 1997,06,01 Set date to June 1, 1997