4 Features and Functions
146 Keysight 34970A/34972A User’s Guide
Front-Panel Operation: First, select the AC current (or AC voltage) function on
the active channel. Then, go to the Advanced menu and select the slow filter
(3 Hz), medium filter (20 Hz), or fast filter (200 Hz) for the active channel. The
default is the medium filter.
LF 3 HZ:SLOW
Remote Interface Operation: Specify the lowest frequency expected in the
input signal on the specified channels. The instrument selects the appropriate
filter based on the frequency you specify (see table on previous page). The
MEASure? and CONFigure commands automatically select the 20 Hz
(medium) filter.
SENS:VOLT:AC:BAND 3,(@203) Select the slow filter (3 Hz)