4 Features and Functions
166 Keysight 34970A/34972A User’s Guide
– Remote Interface Operation: To read the count from the specified totalizer
channel, send the following command. The count may be returned with time
stamp, channel number, and alarm status information depending on the
FORMat:READing command setting (see “Reading Format” on page 118 for
more information).
SENS:TOT:DATA? (@303)
To configure the totalizer reset mode, send either of the following commands
(RRESet means “read and reset”).
SENSe:TOTalize:TYPE {READ|RRESet}[,(@<ch_list>)]
CONFigure:TOTalize {READ|RRESet} ,(@<scan_list>)
To configure the totalizer to count on the falling edge (negative) or rising edge
(positive) of the input signal, send the following command.
SENSe:TOTalize:SLOPe {NEG|POS} ,[(@<ch_list>)]
To immediately clear the count on the specified totalizer channel (whether
scanning or not), send the following command.
SENSe:TOTalize:CLEar:IMMediate [(@<ch_list>)]