4 Features and Functions
168 Keysight 34970A/34972A User’s Guide
– Remote Interface Operation: From the remote interface, you can output an
8-bit byte to one port or a 16-bit word to both ports simultaneously using the
following commands. You must specify a decimal value (binary data is not
accepted). If you are going to read both ports simultaneously, you must send
the command to port 01.
SOUR:DIG:DATA:BYTE 10 ,(@302) Write to port 02
SOUR:DIG:DATA:WORD 10327 ,(@301) Write to both ports